How to use

CellVision slides are easy to use. Each slide consists of 2* seperate chambers, A and B.  The sample volume is indicated in the loading zone*. The grid (if applicable) can be found in the center circle of the chamber. CellVision slides are disposable and each chamber can be used only once.

* except for the 20 micron 4-chamber slide

Filling the chamber

1. Use a pipette to add an appropriate amount of homogeneous sample to one of the loading zones. Do not overfill the chamber.              

               

                        1 = loading zone 2 = grid location

2. The counting chamber will fill itself by capillary action.

3. Wipe away any excess.

4. The sample will stabilize and a steady view will be available immediately.  For the 100 micron chambers we advise to wait 5 minutes before counting. This time is needed for the cells/particles to settle.

We advise these volumes :

Slide depth Volume to add
   
   
  16 µm   4  µl
  20 µm   5  µl
  20 µm-4ch   3  µl
  30 µm   8  µl
100 µm  25  µl

Counting

1. Counting can be performed with the built-in grid.

2. Counting can be performed with a CASA system.

3. Counting can be performed with an eye-piece reticle

Slide depth Built-in grid # of boxes concentration
       
       
20 102 5 106/ml.
100 Neubauer Improved depending on your protocol and dilution  
100 Bürker depending on your protocol and dilution  
100 Bürker-Türk depending on your protocol and dilution  
100 Thoma depending on your protocol and dilution  
       
       

Precautions

When used for analyzing hazardous or potential infectious materials :

* take neccessary precautions

* handle with care

* dispose in an appropriate way

 

 
                                                                                                                                        up     products     homecounting chamber hemacytometer chambers count concentration fertility hygiene accuracy cost-effective boar semen CASA casa eye piece reticle LEJA Leja leja Neubauer improved bürker bürker-türk mallasez fuchs-rosenthal thoma makler nageotte tissue culture blood cell cells sperm spermatozoa buerker buerker-tuerk quality control alternative disposable microscope cleaning glass grid raster patented microcell hamilton thorne research IVOS ultimate cellvue vue cover slip kova hycor mac master pig pigsperm  seminal fluid motility wipe clean tissue life science anthos netherlands holland innovative conception technologies